ID: 372417.0, MPI für Gravitationsphysik / Laser Interferometry & Gravitational Wave Astronomy
Thermal lensing in recycling interferometric gravitational wave detectors
Authors:Strain, Kenneth A.; Danzmann, Karsten; Mizuno, J.; Nelson, P. G.; Rüdiger, Albrecht; Schilling, Roland; Winkler, Walter
Date of Publication (YYYY-MM-DD):1994-10
Title of Journal:Physics Letters A
Volume:194
Issue / Number:1-2
Start Page:124
End Page:132
Review Status:not specified
Audience:Not Specified
Abstract / Description:Thermal lensing limits the performance of advanced interferometric gravitational wave detectors that use high light powers. We evaluate the effects of thermal lensing in such systems and estimate their gravitational wave sensitivity assuming that fused silica optical substrates are employed. Although useful sensitivity can be achieved with established designs, the new technique of resonant sideband extraction is most promising for wideband detectors.
External Publication Status:published
Document Type:Article
Communicated by:Karsten Danzmann
Affiliations:MPI für Gravitationsphysik/Laser Interferometry & Gravitational Wave Astronomy
MPI für Gravitationsphysik/Teilinstitut Hannover
External Affiliations:Max-Planck Institut für Quantenoptik, Hans-Kopfermann-Strasse 1, D-85748, Garching, Germany
Identifiers:ISSN:0375-9601
DOI:10.1016/0375-9601(94)00717-4
Full Text:
You have privileges to view the following file(s):
372417.pdf  [823,00 Kb] [Comment:Online Journal]  
 
The scope and number of records on eDoc is subject to the collection policies defined by each institute - see "info" button in the collection browse view.